Puli Instruments tells you how to maintain the double glass reactor

æœa标题-1.jpg

        The double-layer glass reactor has strict requirements on materials and structure. There are also many issues that need attention in use and maintenance:


        1. Before installation and use, check the instrument carefully, check whether the glass bottle is damaged, whether the interfaces are consistent, and handle it with care during the process.


        2. When wiping each interface, use a soft cloth (can be replaced by napkin), and then apply a little vacuum grease. (Be sure to cover the vacuum grease after use to prevent dust from entering.)


        3. The interfaces of the double-layer glass reactor should not be tightened too tightly, and should be loosened regularly to avoid long-term locking to cause the connector to seize, especially the PTFE switches everywhere should not be tightened excessively, otherwise it will easily damage the glass and stop the machine. The switch should be loosened afterwards.


        4. When using, turn on the power switch first, and then let the machine run from slow to fast. When shutting down, wait until the machine is in a stopped state before turning off the power.


        5. Pay attention to cleaning after use. After each use, you must use a soft cloth to wipe off various oil stains, stains, and solvent remaining on the surface of the machine to keep it clean.


        6. Clean the sealing ring regularly. The method is: remove the sealing ring, check whether there is dirt on the shaft, wipe it with a soft cloth, then apply a little vacuum grease, and reinstall it to keep the shaft and the sealing ring smooth .


        7. The electrical parts must not be exposed to water or damp. When doing any repair or inspection on the glass reactor, be sure to cut off the power and water source first.


        8. The original accessories must be purchased, and other accessories must not be replaced at will, so as not to cause fatal damage to the machine.


        9. When doing any repair or inspection, the glass reactor must first cut off the power and water source.


        10. After the glass reactor is used, the oil stains, stains, and solvent remaining on the surface of the machine must be wiped off with a soft cloth to keep it clean.